3 equipment required, 4 diagnosis steps, 1prepare instruments for diagnostic test – Metrohm 719 S Titrino User Manual

Page 111: 2perform display test

Advertising
background image

4.2.

Diagnosis

719 S Titrino

107

4.2.3 Equipment required:

voltage calibrator,

e.g. 1.773.0010 Metrohm pH Simulator

or 1.767.0010 Calibrated Reference for mV, pH, Ω

µ

S, °C

highly insulated interconnection cable

6.2108.060

cable 3.496.5070
exchange units, if possible with different cylinder volumes (or 3.496.0070 dummy exchange unit)
stop watch or watch with second hand
digital or analogue voltmeter (if need be, connect a calibrated recorder)
2 connecting cables with 4 mm banana plugs
test plug 3.496.8510 (necessary only if plug ‘Remote’ should be checked)
test plug 3.496.8480 (necessary only if plug ‘RS 232’ should be checked)

4.2.4 Diagnosis steps

1

Prepare instruments for diagnostic test

• Power off.

• Disconnect all external connections (cables at rear, except mains cable and key-

board).

• Remove exchange unit.

• Power on and immediately press and hold the <9> key until the power up test pat-

tern disappears.

diagnose press key 0...9

2

Perform display test

• Press <2>.

display test

• Press <Enter>.

Characters for a visual check of the display are generated on the eight lines.

Test sequence:

a) The display is cleared and overwritten from the left to the right with a dot pattern ( ▓ ▓ ▓

▓ ▓ ▓ ▓ ▓ ).

b) The display is cleared and both lines are written into with the letters A, B, C,…Z.

c) The complete character set is shown as a moving display. At the same time with mov-

ing display the LED’s „STATISTICS“ and „SILO“ are swtiches on and off.

• The test sequence can be held and then continued at any time by pressing <5>.

• Block 2 is quit by pressing <Clear>.

diagnose press key 0...9