2.7 Parameters, key <PARAM>

719 S Titrino

27

start V: OFF
cond.




start V 0.0 ml
cond.


factor 0
cond.



dos.rate max. ml/min

**titr.










pause 2 0 s

**titr.




extr.time 0 s

**titr.




meas.input: 1






I(pol) 1 µA

U(pol) 400 mV



electrode test: OFF




temperature 25.0 °C
cond.

Type of start volume (OFF, abs., rel.)

"OFF":

start volume switched off

"abs.":

absolute start volume in mL

"rel.":

relative start volume to sample size.

If "abs." is set:

Absolute start volume (0...999.99 mL)

If "rel." is set:

Factor for relative start volume (0...±999999).

Calculated as: start V (in mL) = factor∗sample size

Dosing rate for start volume

(0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Pause 2 (0...999999 s)

Waiting time after start volume, e.g. reaction time after

dosing of a start volume. The waiting time can be

aborted with <QUIT>.

Extraction time (0...999999 s)

During extraction time no dispensing occurs but the

titration does not stop. The extraction time can be

aborted with <QUIT>.

Measuring input (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 124.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive state to a measurement.

"OFF" means that the test is not performed.

Titration temperature (-170.0...500.0°C).

Temperature is measured at the start of the titration if a

T sensor is connected. The value is used for

temperature compensation in pH titrations.