4.2.

Diagnosis

719 S Titrino

111

Input 2

XX mV

• Carry out the same measurements as with Input 1.

• Press <Clear>.

analog input test 1...5

• Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).

• Press <3>.

Input 1-2

XX mV

The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.

Example: 0 - (+)1500 mV = -1500 mV

• Remove cables from the inputs ‘Ind I’ and ‘Ind II’.

• Press <Clear>.

analog input test 1...5

7.2 To check Pt 100 / 1000 connection

• Connect a Pt 100 or Pt 1000 sensor, a resistor switch-box or a single resistor of 100

Ω or 1 kΩ, respectively, to sockets ‘Pt 100/1000’ by means of short cables.

• Press <4>.

Pt 100*

XX °C

(* or Pt 1000)

Tolerance: ± 0.5 °C (Note also tolerance of resistor switch-box).

The sensor allows automatic displaying of the room temperature. (The resistor values corre-

spond to 0°C.)

• Press <Clear>.

analog input test 1...5

• Remove cables and resistor switch-box.

7.3 Polarizer

test

• Press <5>.

polarizer test

• Press <Enter>.