2.14. Current sample data, key <SMPL DATA>

719 S Titrino

50

2.14 Current sample data, key <SMPL DATA>




The key <SMPL DATA> can be used to enter the

current sample data. The contents of this key change

when the silo memory is switched on, see page 52.

Instead of entering the current sample data with

<SMPL DATA>, you can request these data

automatically after start of determinations.

Configuration: <PARAM>, ">preselections".

Current sample data can be entered live. For working

with the silo memory see page 51.
id#1...3 or C21...C23, sample identifications:

The sample identifications can also be used as sample

specific calculation variables C21...C23.

smpl size:

Sample size.

Entry using keypad or via balance, see page 120.

smpl unit:

Unit of the sample size.

The display texts of the Titrino are shown to the left.
The values are the default values.

smpl data

id#1 or C21

id#2 or C22

id#3 or C23




smpl size 1.0 g


smpl unit: g

Sample data

Sample identification 1...3 or sample specific operand

C21...C23 (up to 8 characters).

Sample identifications or sample specific operands

can be entered using the keypad, via a balance with a

special input device or via barcode reader.

Sample size (6-digit number: ±X.XXXXX)

Entry using keypad, via balance or via barcode reader.

Unit of sample size (g, mg, mL, µL, pc, no unit or up to

5 characters)

Select unit with <SELECT>.

SMPL

DATA

smpl data