2.7. Parameters, key <PARAM>

719 S Titrino

26

max.rate 10.0 ml/min
**titr.










min.rate 25.0 µl/min
**titr.






stop crit: drift
**titr.
stop drift 20 µl/min
**titr.

t(delay) 10 s
**titr.







stop time OFF s
**titr.

Maximum dosing rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

This parameter determines primarily the addition rate

outside the control range, see also page 31.

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Minimum dosing rate (0.01...999.9 µL/min)

This parameter determines the addition rate right at the

start and the end of the titration, see also page 42.

This parameter influences the titration speed and

therefore its accuracy very strongly: A smaller min.rate

results in a slower titration.

Type of stop criteria (drift, time)

Titration stops if EP is and stop drift is reached. (1...999

µL/min)

Titration stops if there is no dosing during t(delay).

(0...999 s, INF)

<CLEAR> sets "INF"

Switch off when the end point is reached and the set

time after the last dispensing has elapsed.

If "INF" is set, an inquiry regarding the stop time

appears.

If t(delay) is "INF"

Stop after a time (0...999999 s, OFF)

<CLEAR> sets "OFF".

Stop after the set time after the start of the titration.

"OFF" means no stop, i.e. titration for an "infinitely" long

time.


>titration parameters

titr.direction: auto









pause 1 0 s

**titr.




General titration parameters

Direction is set automatically (+, -, auto)

auto: The direction is set automatically by the Titrino

(sign [U

first

- EP]).

+:

Direction of higher pH, higher voltage (more

"positive"), larger currents.

- :

Direction of lower pH, lower voltage, smaller

currents.

The titration direction is fixed if two EP's are set. In this

case, an input for titration direction has no meaning.

Pause 1 (0...999999 s)

Waiting time, e.g. for the equilibration of the electrode

after the start. The waiting time can be aborted with

<QUIT>.