9rs 232 test – Metrohm 719 S Titrino User Manual

Page 117

Advertising
background image

4.2.

Diagnosis

719 S Titrino

113

I/O-test-connector?

• Insert the 3.496.8510 test plug in port B 'Remote'. (Do not switch off instrument!)

• Press <ENTER>.

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is con-

nected, error message error 50 01HEX

appears).

2. If no error is found, after about 1 s display shows

extern input/output o.k.

• Remove test plug.

• Press <CLEAR>.

diagnose press key 0...9

9

RS 232 test

A 3.496.8480 test plug normally used in the repair service is required for this test. However, this plug can also be

purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the RS 232 interface is not re-

quired, continue with point 10.

Fig. 4

Connections in the 3.496.8480 plug

• Press <5>.

RS232 test

• Press <ENTER>.

TxD

RxD

DCD

DTR

DSR

RTS

CTS

2

3

8

20

6

4

5