11 data output – Metrohm 719 S Titrino User Manual

Page 48

Advertising
background image

2.11. Data output

719 S Titrino

44

2.11 Data output

Reports for the output at the end of a determination







With <DEF>, the report sequence at the end of the

determination is defined.

The entries are specific to the method and are stored in

the method memory.

report:
Definition of report blocks to be printed automatically at
the end of the determination.


The display texts of the Titrino are shown to the left.
The values are the default values.

>report

report:


report:full

Report sequence: full, short, param, calc, calib, ff

Select a block with <SELECT>. If you require more

than one report block, set a ";" as a delimiter between

the blocks.

Meaning of the report blocks:

full

Full result report with raw results, calculations and statistics.

short

Short result report with calculations and statistics.

param Parameter

report.

calc

Report with formulas and operands.

calib Calibration

data.

ff

Form feed on printer.

Original reports which are put out automatically at the end of the titration can be printed

with recalculated values at any time. Key sequence:

<PRINT><REPORTS><ENTER>.

Original reports have double dashes ==== at the end, whereas recalculations are

marked by single dashes

----

.

Report outputs can be stopped with <QUIT>.

DEF

(

2

def

>report

Advertising