5 configuration, key <config – Metrohm 719 S Titrino User Manual

Page 25

Advertising
background image

2.5 Configuration, key <CONFIG>

719 S Titrino

21

2.5 Configuration, key <CONFIG>

Key <CONFIG> serves to enter device specific data.

The set values apply to all modes.

peripheral units:

Selection of printer, balance and the curve at the ana-

log output.

auxiliaries:

e.g. setting of dialog language, date, time, etc.

RS232 settings:

RS parameters for the COM interface.

common variables:

Values of common variables.

The display texts of the Titrino are shown to the left.

The values are the default values.


>peripheral units

send to: IBM












balance: Sartorius









record: U

Settings for peripheral units

Selection of printer (Epson, Seiko, Citizen, HP, IBM) at

the Titrino Rs232 interface.

"Epson", for Epson

"Seiko", e.g. for DPU-414

"Citizen", e.g. for iDP 562 RS, Custom DP40-S4N

"HP" e.g. for Desk Jet types. Place curves always at the

beginning of a page as you cannot have them over 2

pages.

"IBM" for all printers with IBM character set Table 437

and IBM graphics, as well as for the data transmission

to a computer with Metrodata software TiNET or

VESUV.

Selection of balance (Sartorius, Mettler, Mettler AT,

AND, Precisa)

Sartorius: Models MP8, MC1

Mettler:

Models AM, PM, AX, MX, UMX and

balances with 011, 012, or 016 interface

Mettler AT: Model AT

AND:

Models ER-60, 120, 180, 182, FR-200, 300

and FX-200, 300, 320

Precisa:

Models with RS232C interface

Selection of the curve for the output at the analog

output (U, dU/dt, V, dV/dt, U(rel), T

U:. Voltage

dU/dt: Measured value drift

V: Volume

dV/dt Volume drift

U(rel): Control deviation with SET

T: Temperature with MEAS T

CONFIG

configuration

>peripheral units

Advertising