10 spindle drive and stopcock changeover – Metrohm 719 S Titrino User Manual

Page 118

Advertising
background image

4.2. Diagnosis

719 S Titrino

114

RS232 test-connector?

• Insert the 3.496.8480 test plug in 'RS 232' port.

• Press <Enter>.

Test

sequence:

1. In case of an error an error message is displayed. If for example no test plug is con-

nected, error message

error 68

appears).

2. If no error is found, after about 5 s display shows

RS232 test o.k.

• Remove test plug.

• Press <Clear>.

diagnose press key 0...9

10 Spindle drive and stopcock changeover

• Press <Clear>.

The Titrino fills (only if an exchange unit is inserted)

The Titrino gets out of the diagnosis menu and returns to the user program.

• Remove exchange unit.

• Check spindle zero position, see Fig. 5.

0.2 - 0.6 mm


Fig. 5

Fig. 6

The spindle must be 0.2 - 0.6 mm below the edge of the sliding plate.

The bar of the stopcock coupling must be exactly parallel to the lateral edges of the MPT

Titrino, see Fig.6.

• Reinsert Exchange Unit.

Titrino

fills.

The display of before reappears.

• Knob 'dV/dt' to right stop.

• Press the <DOS> key (on instrument) until the piston rod reaches the top and at the

same time measure the time from start to end.

XXX X

********

cylinder empty !

mind selected language!