2 "tit.hcl, User methods – Metrohm 719 S Titrino User Manual

Page 138

Advertising
background image

6.3. User methods

719 S Titrino

134

6.3.2 "Tit.HCl"

Titer determination of HCl using TRIS.

The titer is calculated as a factor without di-

mension out of 5 determinations and stored

as common variable C36. It can therefore be

used directly by subsequent methods.

Electrode:

6.0232.100 combined glass pH electrode, at

measuring input 1.

Titrant:

c(HCl) = 0.1 mol/L

Sample:

Weigh in Tris(hydroxymethyl)aminomethane

(TRIS) dried 2 h at 105 °C.

Sample size according to the buret volume.

Dilute in 40 mL dist. water.

Reference:

METROHM Application Bulletin Nr. 206:


– Result (Titer) as a digit without dimension

– Sample size TRIS in g

– Theoretical consumption for 1 mol TRIS

– Molar mass of TRIS

C36 is the common variable for the titer

c(HCl) = 0.1 mol/L

'pa

719 S Titrino 18200 719.0021

date 2002-03-13 time 12:24 0

SET pH Tit.HCl

parameters

>SET1

EP at pH 5.10

dynamics 3

max.rate 5 ml/min

min.rate 0.5 µl/min

stop crit: drift

stop drift 20 µl/min

>SET2

EP at pH OFF

>titration parameters

titr.direction: auto

pause 1 0 s

start V: rel.

factor 70

dos.rate max. ml/min

pause 2 5 s

extr.time 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 99.99 ml

filling rate max. ml/min

>statistics

status: ON

mean n= 5

res.tab: original

>preselections

conditioning: OFF

req.ident: OFF

req.smpl size: value

activate pulse: OFF

------------

'fm

719 S Titrino 18200 719.0021

date 2002-03-13 time 12:24 0

SET pH Tit.HCl

>calculations

Titer=C00*C01/C02/EP1;4;

C00= 1.0

C01= 10000

C02= 121.14

------------

'de

719 S Titrino 18200 719.0021

date 2002-03-13 time 12:24

SET pH Tit.HCl

def

>formula

Titer=C00*C01/C02/EP1

RS1 text Titer

RS1 decimal places 4

RS1 unit:

>common variables

C36=MN1

>report

>mean

MN1=RS1

------------

Advertising