3 rack data – Metrohm 789 Robotic Sample Processor XL User Manual

Page 100

Advertising
background image

3.7 Sample racks

92

Metrohm Sample Processor, Operation

3.7.3 Rack

data

In many applications the method and beaker sizes must be observed

exactly. As sample racks are intended for particular beaker sizes, the

rack definitions also contain, in addition to the rack positions them-

selves, data about the lift positions that are directly associated with the

particular beaker size.

The following data can be defined for each rack:

Rack name

Unambiguous identification, ordering

number is standard

Code

Magnet code for automatic rack rec-

ognition

Work position

Working height for Lift 1 and 2

*

Rinse position

Rinsing height for Lift 1 and 2

*

Shift position

Shifting height for Lift 1 and 2

*

Special position

Additional height for Lift 1 and 2

*

Beaker radius

Radius of sample beaker

Beaker sensor

Beaker checking mode

Special beaker positions

16 reserved rack positions with individ-

ual working heights, beaker radius and

beaker sensor test

*

can be set for both towers separately

The rack name is used for the unambiguous identification of a rack.

The standard name stored for the rack is the ordering number. In a

method, this can be assigned to a particular rack name (see Sec-

tion 3.3.1). The automatic rack recognition ensures that if an incorrect

rack is used this will be recognized and the user will be informed of this.

The code is used for automatic rack recognition. You must make sure

that this 6-place binary code coincides with the actual magnet code set

on the rack. Rack codes can be altered as required. However, they

must always be assigned to a single rack only. Avoid the use of the

predefined codes for the standard racks supplied by Metrohm, see also

p. 90.

The working position is used for defining the lift position at which a

sample is to be processed. This means that the ideal setting for a par-

ticular sample rack can be defined depending on the height of the

sample beakers. In manual operation this working position can be

moved to directly with the <END> key. In a run sequence this can be

programmed with

LIFT :1 : work mm

.