3.2

Configuration

Metrohm Sample Processor, Operation

57

3.2 Configuration

CONFIG

Main

menu:

configuration

>auxiliaries

• Open submenu with <ENTER>

configuration

>tower 1

• Use < > or < > to move up and down by one

menu item

configuration

>tower 2

• Use <HOME> or <END> to move to the first or last

menu item

configuration

>rack definitions

• Use <QUIT> to return to operating condition

configuration

>Dosing units

configuration

>RS232 settings


3.2.1 Miscellaneous

configuration

>auxiliaries

Submenu for basic settings

Open the submenu with <ENTER>

>auxiliaries

dialog: english

Select dialog language

english, deutsch,

français,

español


>

auxiliaries

display contrast

3

Set display contrast

0…3…7

0 = low contrast

7 = high contrast

>auxiliaries

beeper: ON

Beep for warnings on or off

ON,OFF

>

auxiliaries

external Start:

off

Switch on external start via Remote line

(Input 7)

on, off

Normally the Sample Processor takes over the complete run control for a
sample series as the "Master". This may be unwanted, particularly if an
automated system has been set up that includes instruments that are not
from Metrohm. This is why the Sample Processor can also be started from
any external instrument via the Remote interface.

With

external Start: ON

the run sequences of the Sample Processor will

Use <QUIT>

for next higher

level