2.4 Connecting up accessories

48

Metrohm Sample Processor, Installation

Fig. 29 Sample rack for XL models

Sample racks for the 789 Robotic Sam-
ple Processor XL (48 cm dia.) have a
handle with fixing screws. This handle is
used to fix the rack on the Sample
Processor turntable by turning it clock-

wise.

2.4.11 Mounting the safety shield

It is essential that the safety shield is mounted before the Sample
Processor is used for the first time. Not only does it provide an
effective protection against splashes, but it also considerably re-

duces the risk of injury.
Various versions of the safety shield are available.

The safety shield is mounted on the Sample Processor tower with
the screws provided. The position of the safety shield can be

modified to suit the height of the sample beakers.

Fig. 30 Mounting the safety shield

Fig. 31 Safety shield for Swing Heads

Attachment screw