1.8 Safety information

24

Metrohm Sample Processor, Introduction

1.8.3 Personal

protection

The different drives of the Sample Processor and the 786 Swing Head

possess considerable force in order that they can freely move even un-

der heavy loads. They are fitted with electronic overload limiters that are

triggered if the mechanical resistance is too high.

When the

instrument is running do not intrude in the working range

of the Sample Processor.
For the user there is a considerable

risk of injury from manual in-

terference with the automatic processes of the instrument.

HOLD

LEARN

Running methods and individual commands can be interrupted with

the <HOLD> key and then continued with <START>.

Emergency stop

STOP

The <STOP> key always terminates all running processes immedi-

ately (exception: initialization during switch-on).


Important!

Before you use the instrument for the first time you must install the

included

6.2751.0xx Safety Shield, see p. 48.

The Sample Processor must not be operated without the Safety

Shield in position!

Fig. 10 Safety Shield

(example shown: 6.2751.0xx for transfer robotic arm)

Biohazard

The Sample Processor does not provide sufficient protection when

processing potentially infectious samples or reagents.

In this case install the necessary safety devices.