5 start-up – Metrohm 789 Robotic Sample Processor XL User Manual

Page 57

Advertising
background image

2.5

Start-up

Metrohm Sample Processor, Installation

49

2.5 Start-up

Some instrument settings must be made on the Sample Processor be-

fore it is used for the first time. These are:

• Dialog language
• Counter setting of running time meter
• Tower configuration
• Robotic arm settings (optional)
• Rack definitions
• Dosing unit configurations
• RS232 interface parameters

All these settings are found in the configuration menu that you can

open with the <CONFIG> key.

Under

>auxiliaries

:

• Set the dialog language to English, see p. 57.

Enter a warning limit under

>>running time meter

, see p. 58.

Recommendation: 1000 h

Under

>tower 1

and

>tower 2

:

Limit the max. stroke path, i.e. set the lowest possible lift posi-

tion, see p. 58. Recommendation: 160…200 mm

(Can be set with the LEARN function)

If a 786 Swing Head is installed then, under

>>swing head 1

, set

the swing position for the robotic arm, select a swing angle for

at least one external position and set its working height, see p.

58ff. All this can be set using the LEARN function.

Under

>rack definitions

:

Load the rack data of the attached sample rack and set the

various lift positions (working position, shifting position, rinsing

position, etc. for each lift) for processing the samples, see p.

61ff. This can all be done by using the LEARN function.

Select a beaker sensor. For single row samples racks the opti-

cal sensor of the tower should be selected.

Under

>>>special positions

you may want to define one or

more special beaker positions. The rack position for the special

beaker (e.g. rinsing beaker, conditioning beaker) must be en-

tered and the working position must be set for each lift (with the

LEARN function), see p. 61ff.

The rack definitions must be saved and then the sample rack

must be initialized with the <RACK> key.