2 pip ext – Metrohm 789 Robotic Sample Processor XL User Manual

Page 138

Advertising
background image

6.2 Standard methods

130

Metrohm Sample Processor, Annex

STIR T1: cont.

STIR T2: cont.

STIR MSB1: cont.

STIR MSB2: cont.

STIR MSB3: cont.

------------

- Do not alter stirrer status

"

"

"

"


6.2.2 PIP

ext

Standard method for pipetting samples into an external titration cell.

Very suitable for standard titrations.

Required instruments and cables:

• Metrohm titrator (Titrino or Titrando with 6.2148.010 Remote Box)

• 786 Swing Head with robotic arm for sample transfer (6.1462.030 or

6.1462.040)

• 803 Magnetic Stirrer or 804 Ti Stand with 802 Rod Stirrer connected

to MSB1

• 800 Dosino with 807 Dosing unit (recommended: 10 mL cylinder,

6.3032.210) and 6.1562.100 Pipetting tubing

• 6.2141.020 Remote cable

Special features:

Samples are pipetted directly from the sample beaker into the external

titration cell. The external Position 1 with the working height for moving

to the titration cell must be set in the tower configuration. No special

beaker is required.

789 Sample Proc. XL 11122 5.789.0011

parameters

method PIP ext

number of samples: rack

>start sequence

1 CTL:Rm: INIT

2 MOVE 1 : ext.1

3 LIFT: 1 : rinse mm

4 PUMP 1.2 : 30 s

5 DOS: 1.1 : prepar. mL

6 PUMP 1.* : 30 s

7 LIFT: 1 : shift mm

>sample sequence

1 MOVE 1 : ext.1

2 LIFT: 1 : work mm

3 PUMP 1.2 : ON s

4 DOS: 1.1 : endVol mL

5 DOS: 1.1 : compen. mL

6 PUMP 1.2 : OFF s

7 LIFT: 1 : shift mm

8 DOS: 1.1 : -0.5 mL

9 MOVE 1 : sample

10 LIFT: 1 : work mm

11 DOS: 1.1 : -5.0 mL

12 WAIT: pause 3 s

13 LIFT: 1 : shift mm

14 DOS: 1.1 : -0.2 mL

15 DOS: 1.2 : port mL

16 MOVE 1 : ext.1

17 STIR: MSB1 : ON s

18 PUMP 1.1 : 10 s

19 LIFT: 1 : work mm

- Report header with device ID and program version

- Method name

- No. of samples (here whole sample rack)

- Initialize Remote interface

- Move robotic arm to external titration cell

- Run lift to rinsing height

- Rinse for 30 seconds

- Prepare Dosing unit

- Rinse titration cell for 30 seconds and aspirate

- Run lift to shifting height
- Move robotic arm to external titration cell

- Run lift to work height

- Rinse cell, pump on

- Eject all dosing cylinder contents

- Eliminate slack

-

Pump off

- Run lift to shifting height

- Aspirate air-gap bubble into pipetting tubing

- Rotate rack to sample beaker

- Run lift to working height

- Aspirate sample into pipetting tubing

- Wait for pressure equilibration in pipetting tubing

- Run lift to shifting height

- Aspirate air into pipetting tip

- Rotate Dosino stopcock, close Port 1

- Move robotic arm to titration cell

- Switch on stirrer

- Add solvent with pump

- Run lift to working height