4 ph_cal – Metrohm 789 Robotic Sample Processor XL User Manual

Page 142

Advertising
background image

6.2 Standard methods

134

Metrohm Sample Processor, Annex

6.2.4 pH_cal

This is a universal method for automated pH measurements directly in

the sample beaker. The electrode calibration is included in the start se-

quence.

Required instruments and cables:

• Metrohm pH Meter (model 691 or 780/781) with 6.2148.010 Remote

box

• Metrohm 802 Propeller Stirrer or 741 Magnetic Stirrer

• 6.2141.020 Remote cable

The method can be used with or without a 786 Swing Head on the

Sample Processor. In the rack definition the special beaker positions 1

to 3 must be reserved.

Special features:

The two buffer solutions for the electrode calibration must be in special

beaker positions 1 and 2.

After each measurement the electrode is rinsed in a rinsing beaker at

special beaker position 3. Internal or external pump required for rinsing.

If a 780/781 pH Meter is used, two additional command lines must be

inserted in the start sequence of the method. This is to cancel the dis-

play of the calibration results:

17 SCN:Rm : end1

18 WAIT: pause 1 s

19 CTL:Rm: START device1

20 MOVE 1 : spec.3

789 Sample Proc. XL 11122 5.789.0011

parameters

method pH cal

number of samples: rack

>start sequence

1 CTL:Rm: INIT

2 MOVE 1 : spec.3

3 LIFT: 1 : work mm

4 PUMP 1.* : 4 s

5 MOVE 1 : spec.1

6 LIFT: 1 : work mm

7 STIR: T1 : 10 s

8 CTL:Rm: METER cal pH

9 SCN:Rm : End1

10 MOVE 1 : spec.3

11 LIFT: 1 : work mm

12 PUMP 1.* : 4 s

13 MOVE 1 : spec.2

14 LIFT: 1 : work mm

15 STIR: T1 : 10 s

16 CTL:Rm: METER enter

17 SCN:Rm : End1

18 MOVE 1 : spec.3

19 LIFT: 1 : work mm

20 PUMP 1.* : 4 s

>sample sequence

1 SHIFTRATE:+ 20

- Report header with device ID and program version

- Method name

- No. of samples (here whole sample rack)

- Initialize Remote interface

- Move to rinse beaker

- Lower electrode

- Rinse electrode and aspirate beaker

- Move to buffer 1

- Immerse electrode

- Stir for 10 seconds

- Start calibration with 1st buffer

- Wait for calibration

- Move to rinse beaker

- Lower electrode

- Rinse electrode and aspirate beaker

- Move to 2nd buffer

- Immerse electrode

- Stir for 10 seconds

- Start calibration with 2nd buffer

- Wait for calibration

- Move to rinse beaker

- Lower electrode

- Rinse electrode and aspirate beaker

- Rack rotation direction counterclockwise