2.3 Data transmission connections

Metrohm Sample Processor, Installation

31

2.3 Data transmission connections

Connection cables
Only Metrohm cables should be used to connect a Metrohm Sample

Processor to other instruments. Only these will guarantee interference-

free data transmission.

Note:

The Remote cables for Metrohm Sample Processors and Sample

Processors have markings at each end of the cable to indicate the in-

strument that the particular connector is intended to be used with and

the connection into which it is to be plugged.

Example:

Titrino B

692 / 712 / 713

Fig. 15 Remote cable

The Sample Processor must be switched off before peripheral devices

are connected, as otherwise the instruments could be damaged.

2.3.1 Remote

connections

Sample Processor — Titrino / Titrando

with standard cable (and 6.2148.010 Remote Box)

B

A

Titrino

C

D

778
789

MSB

778
789

Titrando

6.2148.010

In order to connect Remote cables to Titrandos and the 780/781

pH/Ion Meter an 8.2148.010 Remote Box is required as an additional

adapter; this is connected to one of the MSB sockets (MSB1…4), see

above.

Cable 6.2141.020

Cable 6.2141.020