4 communication commands – Metrohm 789 Robotic Sample Processor XL User Manual

Page 88

Advertising
background image

3.4 Commands

80

Metrohm Sample Processor, Operation

3.4.4 Communication

commands

The coordination of external instruments or the explicit triggering of

functions is possible via both the Remote interface and the serial RS232

interface. The interfaces can be scanned for particular signal patterns or

incoming data strings. In return, individual signal lines can be set or any

character strings (as control commands) can be transmitted to con-

nected instruments.

SCAN

1

SCAN

>sample sequence

7 SCN:Rm

: Ready1

Rm,RS Ready1

Ready2

Ready∗

Cond ok

End1

End2

EndMeter

Continue

8 Bit (1,0 or ∗)

Scans the Remote

interface

= Device 1 ready

= Device 2 ready

= Devices 1+2 ready

= Conditioning finished

= EOD impulse device 1

= EOD impulse device 2

= End impulse 692 Ion Meter/713 pH Meter

= Continue impulse from instrument

any 8-bit binary pattern

In a sequence the SCN:Rm command will halt the run sequence until

the predefined signal pattern has been received.

Predefined binary patterns are available; these can be selected via

simple names (e.g.

Ready1

or

End2

).

Ready

describes a statically set Ready line (Output 0) of a connected

Metrohm instrument.

End

stands for a pulsed signal, e.g. EOD (=End of

Determination, Output line 3). When pulsed signals are being scanned

it is not possible to scan several lines in parallel.

Setting special binary patterns allows the flexible monitoring of con-

nected instruments. The following applies:

0

= line inactive

1

= line active

*

= any line condition

Example:

00000001

= input line 0 is active =

Device 1: Ready1

With the LEARN function the binary patterns (=line conditions) can be

adopted interactively (see p. 113).
Details about the Remote interface can be found in the Technical Ref-

erence.