7 connecting dosing devices and stirrers – Metrohm 789 Robotic Sample Processor XL User Manual

Page 38

Advertising
background image

2.2 Instrument setup

30

Metrohm Sample Processor, Installation

2.2.7 Connecting dosing devices and stirrers

Metrohm Sample Processors have various connections to which dosing

devices and stirrers can be connected.

Stirrers

The 722/802 Rod Stirrers and 741 Magnetic Stirrer have a DIN
connector and are connected to the rear panel of a tower (Tower stirrer
T1 and T2).

As with other Metrohm instruments, e. g. Titrandos, the three MSB

connectors (MSB = Metrohm Serial Bus) are very versatile. You can

connect an 801 Stirrer or an 804 Ti Stand to them.

Dosing devices
Up to three 700/800 Dosino or 685/805 Dosimat type dosing drives

can be operated by connecting them directly to the MSB sockets of a

Sample Processor. One 800 Dosino or 805 Dosimat can also be con-

nected to the MSB output connector of another instrument connected

to the MSB socket, such as the 801 Stirrer, in a so-called 'daisy chain'.

Please note that only one instrument of a particular type may be pre-

sent in such a daisy chain, e. g. a 801 Stirrer and a 805 Dosimat.
A 700 Dosino or a 685 Dosimat must be connected directly to the

MSB socket of the Sample Processor.

R S 2 3 2

M S B 1

M S B 3

M S B 2

R e m o te

Fig. 14 MSB connections

The instrument will only recognize automatically a connected dosing
device when it is switched on or after RESET.

Dosing device and

stirrer connections

MSB1…3