3.9 The Remote interface

Metrohm Sample Processor, Operation

111

METER Mode I

*********0100*

switches ion meter or pH meter to Ipol (mV measure-
ment) and starts it

METER Mode C

*********1000*

switches ion meter to Conc measurement

METER Cal pH

*********0101*

switches ion meter or pH meter to pH calibration and
starts it

METER Cal C

*********1001*

switches ion meter to Conc calibration

METER enter

*********1111*

simulates the <ENTER> key for ion meter or pH meter

(essential for pH calibration with 691/780/481, in order to
start the measurement of the second buffer)


For

START

commands the signal is transmitted as a short impulse of 200 ms.


*

)

With pH meters or ion meters a result report is triggered

3.9.5 Manual stop options
Parameter Signal

pattern

Function

STOP Device 1

************1*

stops device 1

(e.g. Titrino, Titrando...)

STOP Device 2

******1*******

stops device 2

(see above, only with multiple cable)

STOP Device *

******1*****1*

stops devices 1 and 2 "


For STOP commands the signal is transmitted as a short impulse of 200 ms.