3.9 The Remote interface

110

Metrohm Sample Processor, Operation

Input lines that are not of interest or for which no defined condition can

be predicted should also be masked with an asterisk (∗).

If

external START

(see configuration 3.2.1) is switched on then Re-

mote lines Input 7 and Input 6 will be reserved for external START and

STOP.

In order to simplify the use of Remote commands, particularly when

several devices are linked together with Metrohm cables, predefined bi-

nary patterns are available as command parameters for the commands

CTL and SCN for standard conditions (1 to 2 Titrinos, possibly with

auxiliary Dosimats, pH meter or ion meter). These are:

3.9.3 SCN

command

Parameter Signal

pattern

Function

Ready1

*******1

waits for "ready" condition of device 1 (Titrino, Titrando)

Ready2

**1*****

waits for "ready" condition of device 2

Ready*

**1****1

waits for "ready" condition of devices 1 and 2

Cond ok

******1*

waits for "cond ok" condition of device 1

End1

****1***

waits for the EOD impulse from device 1 (Titrino, Ti-

trando)

End2

*1******

waits for the EOD impulse from device 2

EndMeter

***11***

waits for the EOD impulse from ion meter or pH-Meter

(stirrer 1 will be switched on during the waiting period)

Continue

***1****

waits for an impulse to continue e. g. "Sample ready"

With the parameter

Ready*

it is possible to scan the condition of de-

vices working in parallel. In this case the Ready line of both devices

must be statically set at the end of a determination. Devices which only

transmit a brief impulse at the end of e.g. a measurement cannot be

controlled in parallel.

3.9.4 CTL

command

Parameter Signal

pattern

Function

INIT

00000000000000

initializes the Remote interface

START device1

*************1

starts device

1

(e.g. Titrino, Titrando, ...) *

)

START device2

********1*****

starts device 2

(see above, only with multiple cable)*

)

START device*

********1****1

starts devices 1 and 2 " *

)

SAMPLE ready

*********1****

sends a pulse for continuing a determination

START Dos1

*******1******

starts Dosimat at device 1 (Titrino via "activate")

START Dos2

*****1********

starts Dosimat at device 2 "

START Dos*

*****1*1******

starts Dosimat at devices 1 and 2 "

METER Mode pH

*********0001*

switches ion meter or pH meter to pH measurement and
starts it

METER Mode T

*********0010*

switches ion meter or pH meter to temperature meas-
urement and starts it

METER Mode U

*********0011*

switches ion meter or pH meter to mV measurement and
starts it