3.8 Dosing and liquid handling

106

Metrohm Sample Processor, Operation

3.8.8 Preparing the Dosing unit

The Dosing unit must first be prepared, i.e. the filling tubing and pipet-

ting tubing must be rinsed and completely filled. The cylinder contents

must then be ejected.

Command sequence:

MOVE 1 : ext.1

LIFT: 1 : work mm

DOS: 1.1 : prepar. mL

DOS: 1.1 : EndVol mL

DOS: 1.1 : compen. mL

11


Important!
Eject the cylinder con-
tents

This step should be carried out before every sample series involving pi-

petting. Before each pipetting process the Dosing unit cylinder contents

must be completely ejected. This is the only way to achieve reproduci-

ble, accurate pipetting.

The titration cell can then be aspirated, rinsed and prepared with fresh

solvent.

3.8.9 Pipetting

The pipetting procedure takes place in five steps:

• Ejection of cylinder contents (see above) and formation of separat-

ing bubble

• Move to sample

• Aspirate / Measure out sample

• Move to target

• Eject sample