3.2 Configuration

64

Metrohm Sample Processor, Operation

3.2.4 Dosing

units

Metrohm Exchange units (with 685 and 805 Dosimats as the dosing drive) and Dosing units for
the Dosino 700 and 800 systems can be used with the Sample Processors. The following set-
tings are used for preparing the Dosing units (PREP function). The tubing dimensions (length
and diameter) are used for calculating the rinsing volumes.
If Exchange units are used then only the dosing rate is effective.

configuration

>Dosing units

Submenu for the Dosing unit settings

Open the submenu with <ENTER>

>Dosing units

Dosing unit

1

Selection of Dosing unit or dosing device
connection

1…3

The Dosing unit selection must be confirmed with <ENTER>. It will then be
shown in the first menu line.

>Dosing units

1

max. rate 1

160 ml/min

Max. dosing and filling rate for Port 1

(depends on cylinder size)

0.01…160 ml/min

>Dosing units

1

tube length 1

1000 mm

Length of tubing at Dosino Port 1

0…1000…30000 mm

>Dosing units

1

tube diameter 1

2 mm

Diameter of tubing at Dosino Port 1

0.1…2…20 mm

>Dosing units

1

max. rate 2

160 ml/min

Max. dosing and filling rate for Port 2
(depends on cylinder size)

0.01…160 ml/min

>Dosing units

1

........... to port 4

Input of tubing parameters for all four
ports of a Dosing unit.

Use <QUIT>