3.4

Commands

Metrohm Sample Processor, Operation

77

LIFT

9

LIFT

>sample sequence

3 LIFT: 1:

rest mm

1,2,∗

work,

rinse, shift,

special,

rest,

0…235

mm

Lift positioning

1

st

parameter: tower

2

nd

parameter: position

Raises or lowers one or both (∗) lifts to a defined position.

Examples:

LIFT: 1: work

LIFT: 1: rinse

LIFT: 2: shift

LIFT: 2: special

LIFT: *: rest

LIFT: 1: 100 mm

Moves lift at Tower 1 to working position

Moves lift at Tower 1 to rinsing position

Moves lift at Tower 2 to shifting position

Moves lift at Tower 2 to special position

Moves both lifts to uppermost position (0 mm)

Moves lift at Tower 1 to position 100 mm

Working, rinsing, shifting and special positions are defined specifically

for each rack in the configuration menu under

>rack definitions

(see

p. 61). The rest position is the zero position (0 mm) of the particular lift,

i.e. its upper stop.
The lift speed can be set in the parameter menu under

>sample

changer settings

or altered in a sequence with the appropriate DEF

command.
Each lift can be positioned with an accuracy of one millimeter. The

LEARN function is available for this (see p. 112).

3.4.2 Switching components

Pumps and stirrers can be switched as required, either together or

separately. They work independently and can be operated at the same

time as other functions are being carried out.

PUMP

PUMP

4

>sample sequence

4 PUMP 1.1

: 1 s

1.1…2.2

1…999 s,

1.∗,2.∗ ON,OFF

Pump control

1

st

parameter: pump selection

2

nd

parameter: status or duration


With the PUMP command up to 4 pumps (2 pumps per tower) can be

controlled separately. The first parameter selects the pump.
Syntax: T.P (T=Tower, P=Pump)
With Sample Processor models that have external pump connections

instead of built-in pumps the PUMP command switches the appropri-

ate pump connections (16 V).
The pump built into a tower as Pump 1 is always used for rinsing the ti-

tration head. Pump 2 (if present) can be used for aspirating off the

Advertising