1.7 The keypad

14

Metrohm Sample Processor, Introduction

hand side of the keypad are used for navigation in the menus or for al-

tering parameters. For entering parameters the numerical block on the

right-hand side of the keypad is also available.

The lowest row of keys (<HOLD>, <STOP>, <START>) are used

for the direct control of a method sequence.

1.7.1 The

display

The display consists of two lines each with 24 characters.

The first line is used as the title line, in which the current method and

the sample counter count are shown. In the editing mode the menu title

appears here.

The second line is used as the status line, which shows specific activi-

ties depending on the operating status. In the editing mode it is used as

the input line.

Normal condition

Sample

counter

Method name

Pump status

******** counter 1/12
PUMP--++STIR+---- ready

Instrument status

Stirrer status


Method sequence

Running sequence

******** counter 2/12
START 03 WAIT 11 s

Parameter

Current command with line number


Editing mode

Menu title

Menu line or command

>Sample sequence
1 MOVE 1 : Sample

2

nd

parameter

1

st

parameter

If the Sample Processor is included in a computer-controlled automa-

tion system and is completely controlled via the RS232 interface then it

may be advisable to switch off the display. This can be done in the

setup menu of the Sample Processor, see p. 114.