5.2 Initializing the working memory

Metrohm Sample Processor, GLP validation – diagnosis

123

5.2 Initializing the working memory

With this diagnostic step the default values for instrument parameters

can be entered via the keypad and the instrument returned to its origi-

nal condition. This measure is important with respect to the following:

In rare cases it could happen that massive interference signals such

as current peaks, lightning, etc. could affect the contents of the data

memory. If the data memory contents are undefined then this could

lead to a system crash.


The Sample Processor has different ways of initializing the working

memory. Either the complete data memory (

all

) or only parts of it

(pa-

ram, config, setup, assembly

) can be overwritten with the default

values.

Although the instrument number is retained, initialization should only

be carried out when absolutely necessary, as this deletes the stored

user data (etc.).

• Keep key <9> pressed down when the instrument is switched on.

diagnosis

>RAM initialization

• Press the <ENTER> key to open the following diagnosis menu:

>RAM initialization
select: param

>RAM initialization
select: config

>RAM initialization
select: setup

>RAM initialization
select: assembly

>RAM initialization
select: all

Set method parameters with default values.

Set instrument configuration with default values.

Set setup parameters with default values.

Set assembly parameters with default values.

Delete all user-defined methods

The submenus are selected in sequence by pressing the <Select>

key. The individual initialization versions are accessed by the <EN-

TER> key and exited with the <QUIT> key.