2.2 Instrument setup

28

Metrohm Sample Processor, Installation

Swing

Head


Connecting to the Swing Head connection socket
Lead the connection cable of the 786 Swing Head through the guide
chain of the tower and connect the mini-DIN plug of the Swing Head

connection cable to the Swing Head socket on the rear panel of the
tower.

Configuring the Swing Head and robotic arm

As each type of robotic arm has its own dimensions, it is essential to

adapt the Swing Head setup settings to suit the particular type of ro-

botic arm.

The necessary alterations are made in the setup dialog, see Section

3.13. The individual settings affect:

• Swing offset (default: )

• Max. swing angle (default: 84°)

• Swing radius (=length of robotic arm, default: 110 mm)

• Swing direction (default Tower1: , default Tower 2: +)

The setup dialog of the Sample Processor is opened by pressing the

<CONFIG> key as the instrument is switched on. A short explanation

is given in Section 3.13. Details about the individual settings and types

of robotic arm can be found in the Instructions for Use of the 786 Swing

Head.

2.2.5 Mounting the robotic arm

Certain robotic arms can, depending on the model, be mounted so that

they swing to the right or the left. During the mounting process the loca-

tion of the limiting screw of the robotic arm must be taken into account.

Fig. 12 Robotic arm with limiting screw

Swing-right mounting means: The robotic arm can swing from the

zero-axis (center of the rack) to the right, as seen from the front. The

swing direction has to be set to '

'.

Limiting screw
(for swing-right mounting)

Limiting screw

(for swing-left mounting)