3.8 Dosing and liquid handling

Metrohm Sample Processor, Operation

105

3.8.6 Pipetting

equipment

Together with an 700/800 Dosino as dosing drive (or "pipetting pump"),

the Sample Processor is exceedingly suitable for pipetting liquids in a

volume range from 0.1 to 10 mL.

Equipment required

• Sample Processor with 786 Swing Head
• Robotic arm with transfer head (e.g. 6.1462.030 or 6.1462.040)
• 800 Dosino with 2 mL, 5 mL, 10 or 20 mL Dosing unit
• Pipetting tubing 6.1562.020 (3 mL) or 6.1562.100 (10 mL) with

drawn out tip

• Any sample rack
• Possibly Stand Support 6.2001.070 with titration cell

3.8.7 Pipetting

procedure

Basic principles

Measuring out the liquid to be pipetted takes place in a pipetting tubing

filled with a carrier liquid (preferably water). In order to prevent any

carryover or mixing an air bubble (air gap) must be inserted between

the sample and the carrier liquid. The tubing tip should be drawn out by

hand to form a fine point.

A Dosino drive is used to transfer the liquids into the pipetting tubing.

The pipetting tubing is connected to Port 1 of the Dosing unit. The Dos-

ing unit is mounted on a reagent bottle containing the carrier liquid. This

can be refilled via filling port 2 of the Dosing unit. The carrier liquid can

also be used for rinsing and diluting the sample.

Various pipetting procedures are possible. The following conditions

must always be observed:

• The sample must always be measured out in the pipetting tubing; it

must never enter the Dosing unit cylinder.

• The sample should always be aspirated after the dosing cylinder

has been ejected.

• The difference in level between the surface of the sample liquid and

that of the carrier liquid in the cylinder should be as small as possi-

ble.

• Aqueous sample solutions can be pipetted without having to rinse

the tubing. Non-aqueous samples require the pipetting tubing to be

rinsed with carrier liquid (same solvent as the sample), to avoid

carryover.