3.3 Composition of a method

70

Metrohm Sample Processor, Operation

3.3.4 Dosing device settings
Both Dosinos (models 700 and 800, with Dosing units) and Dosimats (models 685 and 805,
with exchange units) can be used as dosing drives on a Sample Processor. However, free as-
signment of the ports for the various functions is only possible with Dosinos.

parameters

>Dosing unit def.

Submenu for setting the Dosing units

Open the submenu with <ENTER>

>Dosing unit def.

dosing drive

1

Selection of the dosing device, or the

connection

1…3


After entering the connection of the dosing device (see p. 8) and confirming

its with <ENTER> the settings for the selected dosing device should then
be made.

>dosing drive

1

dos.rate max.

mL/min

Set the dosing rate

0.01…160

mL/min,

max.

The maximum possible dosing rate depends on the size of the dosing cylin-

der (rule of thumb: cylinder volume x 3.3).

>dosing drive

1

fill.rate max.

mL/min

Set the filling rate

0.01…160

mL/min,

max.

The maximum possible filling rate depends on the size of the dosing cylinder
(rule of thumb: cylinder volume x 3.3).

The following entries only apply to 700/800 Dosinos. Details about Dosinos
and Dosing units can be found on p. 95ff.

>dosing drive

1

dosing

port 1

Define standard dosing port 1

1…4

>dosing drive

1

dosing 2

port 3

Define standard dosing port 2

1…4

>dosing drive

1

filling port

2

Define standard filling port

1…2…4

>dosing drive

1

rinsing port

2

Define standard rinsing inlet (if Dosing
unit is exchanged)

1…2…4

>dosing drive

1

Define standard outlet for the prepara-