5 rs232 settings – Metrohm 789 Robotic Sample Processor XL User Manual

Page 73

Advertising
background image

3.2

Configuration

Metrohm Sample Processor, Operation

65

3.2.5 RS232

settings

The correct settings of the interface parameters of the serial RS232 interface is essential for the
perfect functioning of data transmission to and from the Sample Processor. This includes the
control of the instrument by using a PC software such as TiNet or Tiamo, backing up of con-
figurations or methods with the VESUV software from Metrohm or printing out a parameter re-
port on a connected printer.

configuration

>RS232 settings

Submenu for serial interface settings

Open the submenu with <ENTER>

>RS232 settings

baud rate:

9600

Transmission rate
in baud

300,600,1200,

2400,4800,9600, 19200

>RS232 settings

data bit:

8

Number of data bits

7,8

>RS232 settings

stop bit:

1

Number of stop bits

1,2

>RS232 settings

parity: none

Parity selection

even, odd,

none

>RS232 settings

handshake: HWs

Handshake selection

HWs,SWchar,SWline,

none

>RS232 settings

character set:

IBM

Character set for printer and PC
(printer emulation)

IBM,HP,Epson,

Seiko,Citizen

The settings for the printers recommended by Metrohm are listed on p. 37ff.
If your printer is not shown there then we recommend the setting "Epson". It
may be necessary to consult the printer manual. For data transmission using
personal computers you should select "IBM".

>RS232 settings

RS control:

ON

Switch data reception on/off

ON,OFF


If Remote control is switched off then no data will be received, but reports
can still be printed out.

Use <QUIT>

for next higher

level

Advertising
This manual is related to the following products: