2 sample processor settings – Metrohm 789 Robotic Sample Processor XL User Manual

Page 76

Advertising
background image

3.3 Composition of a method

68

Metrohm Sample Processor, Operation

3.3.2 Sample Processor settings

parameters

>changer settings

Submenu for setting the Sample Processor functions

Open the submenu with <ENTER>

>changer settings

rack name:

*

The sample rack assigned to the method

*,10 ASCII characters

* = no particular rack

With this setting the user can be forced to use a particular rack for the se-
lected method. If this is not required then

rack name: *

must be selected.

>changer settings

lift rate T1

25 mm/s

Vertical movement rate of Lift 1

5…25 mm/s

>changer settings

lift rate T2

25 mm/s

Vertical movement rate of Lift 2
(only for 2-tower models)

5…25 mm/s

>changer settings

shift rate

20°/s

Speed at which the rack rotates

in degrees per second

3…20

>changer settings

shift direction:

auto.

Direction of rotation of the sample rack

+,–,auto. auto. = the Sample Processor selects the shortest

path for the rotation itself.

>changer settings

rotat. increment:

5.00°

Increment angle for the command

MOVE +/-rotate

0.00…5.00…270.00°

>changer settings

swing rate T1

55°/s

Swing rate of the robotic arm at Tower 1
in degrees/second

10…55

>changer settings

swing rate T2

55°/s

Swing rate of the robotic arm at Tower 2

in degrees/second

10…55

(only for 2-tower models)

>changer settings

swing increment

10.00°

Increment angle for the command
MOVE +/-swing

0.00…10.00…180


>changer settings

on beaker error:

MOVE

Definition of reaction if beaker is missing
(only applicable with beaker sensor on)

MOVE, Message

MOVE = the last action will be carried out

again. The next position according to

the current SAMPLE command will

be selected.

Message = The run will be interrupted and a

warning displayed.