8-20 ADS FlowShark Pulse O&M Manual

Sensor

No.

Air-Ultrasonic

01

Depth measurement using an air-ultrasonic sensor
connected directly to the FlowShark Pulse.

Requires an air-ultrasonic sensor.

Water-Ultrasonic

02

Standard version. Flow velocity and depth
measurement using combi sensor from the bottom
up.

Requires a 2K combi sensor.

External Sensor

03

Second standard version. Depth measurement using
an external, 2-wire sensor powered by FlowShark
Pulse or by external measurement unit and depth
input via mA input signal.

Requires an external 4-20 mA sensor.

Fixed Value

04

Use this point to set parameters for pipes and pipes
that remain full. These applications typically do not
require depth measurements. Enter the consistent
depth level under Scale/Height.

Pressure

05

Depth measurement determined by a sensor-
integrated pressure measurement cell.

Requires a 3K or 4K combi sensor.

Pressure + Air-
Ultrasonic

06

Depth measurements performed using a sensor-
integrated pressure measurement cell and an air-
ultrasonic sensor connected directly to a FlowShark
Pulse.

Requires a 3K or 4K combi sensor and an air-
ultrasonic sensor.

Pressure + Water-
Ultrasonic

07

Depth measurements performed using combination
of sensor-integrated pressure measurement cell and
water-ultrasonic from bottom up.

Requires a 3K or 4K combi sensor.

Pressure + External
Sensor

08

Depth measurements performed using combination
of sensor-integrated pressure measurement cell and
external 2-wire sensor powered by FlowShark Pulse
or by an external measurement unit and depth input
via mA input signal.

Requires a 3K or 4K combi sensor and an external
4-20 mA sensor.