Transmitter installation and connection, General, 2 ads flowshark pulse o&m manual – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 47: Corrosive chemicals or gas, Mechanical shock, Walkways or heavy traffic areas, Vibration

Advertising
background image

6-2

ADS FlowShark Pulse O&M Manual

Transmitter Installation and Connection

General

When selecting a mounting location for the transmitter, avoid locations exhibiting or
near the following conditions:

z

Direct sunlight (use weatherproof cover, if necessary)

z

Heat-emitting objects (maximum ambient temperature of 104°F (40°C))

z

Objects generating strong electromagnetic fields (such as frequency
converters or electric motors that consume significant power)

z

Corrosive chemicals or gas

z

Mechanical shock

z

Walkways or heavy traffic areas

z

Vibration

z

Radioactive emissions

To fasten the wall mount enclosure, use one of the following fastener configurations
based on the composition of the mounting structure:

z

Four M5 machine screws of suitable length with corresponding nuts and
shims. These screws must penetrate the wall at least 1.575 inches (40 mm).

z

Four wood screws with a minimum diameter of 0.177 inches (4.5 mm).
These screws must penetrate the appropriate dowels at least 1.97 inches (50
mm).

The clear door on the measurement transmitter has a protective film to prevent
potential damage during transport and scratches during assembly. Remove this film
immediately after assembly.

Exposing the film to direct solar radiation for a long period will
make it more difficult to remove the film.

Clean the door with mineral spirits or car polish, when necessary. If these cleansers
are not effective, order a new front door from ADS.

The FlowShark Pulse Model 10/20 transmitter does not support
passive OCS sensors (older models).