6-22 ADS FlowShark Pulse O&M Manual

Connect the sensor cable to the transmitter at the termination block. Refer to the
following diagram to connect a flow velocity or water-ultrasonic combi sensor:

not occupied

G1

G2

G3

G4

G5

G6

G7

not occupied

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

white

green

blue

red

black (shield, no earth)

8.6 V

LIYC 11Y 2 1.5mm

.

2

+1 2 0.34mm

.

2

.

max. 492.1 ft

combi sensor supporting flow velocity

or water-ultrasonic measurement

Figure 6-27 Connecting a combi sensor supporting flow velocity or water-ultrasonic

not occupied

F1

F2

F3

F4

F5

F6

F7

not occupied

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

white

green

blue

red

black (shield, no earth)

8.6 V

LIYC 11Y 2 1.5mm

.

2

+1 2 0.34mm

.

2

.

max. 492.1 ft

combi sensor supporting

flow velocity measurement

Figure 6-28 Connecting a second velocity sensor to the FlowShark Pulse Model 20

not occupied

E1

E2

E3

E4

E5

E6

E7

not occupied

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

white

green

blue

red

black (shield, no earth)

8.6 V

LIYC 11Y 2 1.5mm

.

2

+1 2 0.34mm

.

2

.

max. 492.1 ft

combi sensor supporting

flow velocity measurement

Figure 6-29 Connecting a third velocity sensor to the FlowShark Pulse Model 20

not occupied

G1

G2

G3

G4

G5

G6

G7

not occupied

outer shield

supply +

UE-GND

RxTx -

RxTx +

FlowShark Pulse

(white)

(green)

(blue)

(red)

(black; shield, no earth)

8.6 V

A2Y (L)2Y

or similar

max 820.2 ft

connection box with air pressure compensation

(pressure compensation element)

LIYC 11Y 2 1.5mm

.

2

+1 2 0.34mm

.

2

.

+PA 1.5/2.5

max. 98.4 ft

combi sensor supporting flow

velocity & pressure measurement

Figure 6-30 Connecting a flow velocity sensor with an integrated pressure measurement cell