Troubleshooting – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 178

Advertising
background image

C H A P T E R 11

Troubleshooting

Error

Possible Reason

Correction

Connections

Check the connection between the sensor
cable and terminal strip. Check the
connection boxes, sensor extension cable
connections, and air compensation element
for proper connection or moisture.

Check sensor installation for proper
orientation toward the flow and horizontal
position.

Sensor

Check the sensor for existing soil,
sedimentation, or silt, and remove
obstructive debris as necessary.

Check the sensor for mechanical, body, or
cable damage, and replace the sensor,
when necessary.

Important: When no flow depth exists,
measuring flow velocity is not possible.

Water-ultrasonic measurement: Check the
sensor for horizontal installation. Check for
sensor function in I/O > Sensors > H-
Sensor > echo profile
menu.

External level measurement: Check the
external transmitter for function and signal
transmission. Inspect cables, clamp
connections, short circuits, and contact
resistances.

Pressure cell measurement: Inspect the
compensation tube and hardware for
obstructions.

No Indication of Flow
(0 or
----)

Flow Depth
Measurement

When flow depth is greater than 2.56 inches
(65 mm), the FlowShark Pulse is in Q/H
measurement mode at initial start-up.
Manually enter the velocity at a depth of 2.56
inches (65 mm) in the CAL > Flow vel. >
Velocity h_crit
parameter. Perform this
procedure only one time.