8-26 ADS FlowShark Pulse O&M Manual

Connecting 2 or 3 sensors: Selecting 2 or 3 flow velocity sensors displays the
following screen:

Figure 8-36 Extended sensor settings

Select the sensor corresponding to Flow Velocity/Sensor No. The following screen
displays.

Figure 8-37 Selecting sensor type and mounting position

Sensor Type

Selecting and choosing the parameters for several flow velocity sensor types involves
the same procedure used for only one velocity sensor.

Mounting Place

Enter the mounting height for each sensor on this screen. Use the lowest sensor
(must be sensor 1) as the point of reference for determining the height of the other
sensors. (refer to Figure 8-38).

Installed at the lowest point in the channel or pipe, sensor 1 serves
as the guide for the height settings for the other flow velocity
sensors.

If several flow velocity sensors are in use and one of them functions
in conjunction with ultrasonic depth measurement from the bottom
up (water-ultrasonic) or with the pressure measurement cell,
designate this sensor as sensor 1 within the sequence and install it
at the lowest point of the installation.