Air-ultrasonic sensors, Installation 6-15 – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 60

Advertising
background image

Installation 6-15

1

Cutting ring inside and outside

2 Thread
3 Inner

cone

4

Inner thread of spigot nut

Figure 6-18 Greasing parts

Air-Ultrasonic Sensors

Install the air-ultrasonic sensor with clamping using an pipe mounting system. To
install the sensor using this mounting system, run the mounting sheet located in the
pipe vertex (crown) through the cut-out (Item 4 on Figure 6-19) in the air-ultrasonic
sensor prior to final assembly.

1

Ground Plate 1

2

Ground Plate 2

3

Ground Plate 3

4

Cut-out for Pipe Mounting Plate

Figure 6-19 Installing the air-ultrasonic sensor using the pipe mounting system

Before clamping the system into the pipe, adjust the sensor so that it is completely
parallel to the water surface. When installing an air-ultrasonic sensor and combi
sensor measuring flow velocity together, locate the air-ultrasonic sensor upstream at
least 4 inches (10 cm) ahead of the combi sensor at the bottom.