Channel number, Name, Function – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 121

Advertising
background image

Parameter Settings

8-31

This section enables the user to set and assign the digital input signals Control
OPEN
, Control CLOSED, and Torque CLOSED. The FlowShark Pulse Model 20
requires these input classifications for regulator operation.

The Lock v-measurement function is available only for digital input 1.

Figure 8-44 Digital input functions

Locking measurements may be useful for challenging applications, such as locations
that characteristically reflect backwater conditions. The unit either releases or locks
the velocity measurement based on a limit value contact, such as a discrete echo
sounder, float switch, or conductive transducer, using the overflow height as the
switching threshold. Locking refers to setting the flow velocity readings to 0. Since
the unit will calculate volume as V=0, no analog or digital values corresponding to
flow will be available.

However, selecting the option to save individual gates will ensure individual velocity
readings are saved.

Channel number

Select the appropriate digital input, from inputs 1–4, for which to set the following
parameters.

Name

(optional) Enter the name to which to save the digital input to the memory card. This
card is the only location in which this name will be saved.

To enter the name, follow the same procedure used in PAR > Measurement Place >
Name
.

Function

Assign regulator setting functions to the digital input selected under the Channel
Number
. Select from among the following functions by pressing the ALT key:

z

Disabled

z

Control CLOSED Routes the slide end switch for closed condition to the
selected digital input