Parameter Settings

8-61

Figure 8-99 Basic selection menu

The subsequent submenu H-Sensor(s) may vary based on the depth sensor selected:

Figure 8-100 Menu with water-ultrasonic sensor from bottom up

V-Sensor

Selecting this point displays a 2-page table that includes all the individual velocity
readings and the heights of the respective measurement windows.

Figure 8-101 Individual velocity readings

+

Toggle between both pages (measurement windows or gates 1-8 and 9-16) using the
up and down arrow keys.

The absence of a reading (------ ) in a measurement window indicates that flow
velocity currently is not available for measurement in the associated window. This
may occur when the water is very clear or vorticity exists within the area. This also
may occur with low flow depths, such as approximately 13.78 inches (35 cm), where