Use in accordance with the requirements, Ex-protection, Approval – ADS Environmental Services FlowShark Pulse QR 775004 A1 User Manual

Page 19: Electronic specifications

Advertising
background image

2-2

ADS FlowShark Pulse O&M Manual

Use in Accordance with the Requirements

The ADS FlowShark Pulse and corresponding sensors are designed for measuring
slightly- to heavily-polluted flow in partially-filled and full pipes or other similar
applications. Please adhere to the measurement specifications designated in this
chapter. Monitoring flow under conditions falling outside the documented
specifications of the equipments’ capabilities without prior written permission of
ADS occurs at the owner’s risk.

Modifying or using this equipment for purposes other than the
intended use (described above) without the prior written consent of
the manufacturer will be considered outside the specifications of
the equipment. Therefore, any resulting damages that may occur
will be at the user’s risk or expense.

This equipment has an estimated lifetime of 10 years. Therefore,
inspect and overhaul equipment exceeding 10 years in age.

Ex-Protection

The Ex-version of the FlowShark Pulse active sensor is designed for use in explosive
environments (zone 1).

Always install the transmitter outside Ex-zones!

Approval

Sensor:

II 2 G EEx ib IIB T4

Transmitter:

II(2)G [EEx ib] IIB

Electronic Specifications

Analog sensor connections:

Ignition protection type intrinsically-safe

EEx ia IIB

Clamps D8, D9

for connecting certified sensors only

Max values:

U

0

= 23.1 V

I

0

= 162 mA

linear

characteristic

Advertising