4 default indicator, Default indicator – Rice Lake CLS-420 Cargo Lift Scale Installation Manual User Manual
Page 29
CLS-420 Installation Manual
10
4.1.4
Default Indicator
Figure 4-5. Reset Switch
•
If the module setting corrupts or if the indicator is moved to a new facility, and there is a problem
getting it to connect to either the wired, or wireless network, a factory default switch is provided to
wipe ALL configuration in the Matchport to a known state.
Note
This function will reset ALL settings, including the Wireless and IP settings and will set the module up for Wired
Connections ONLY.
Setting the defaults with the switch
1. Remove the small screw from the upper right corner of the back panel.
2. With the indicator off, press and hold the switch.
3. Turn the indicator on and wait 10 seconds,
4. Turn the indicator off and release the switch.
The module is now set to factory defaults and in Wired Mode only.
Note
Reconfigure by connecting the indicator to a Wired Ethernet network using the DeviceInstaller or browser.
When complete, make sure to set the network mode back to “Wireless Only”.