8.0 Appendix

8.1

Error Messages

The 420 indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator LED display.

8.1.1

Displayed Error Messages

The 420 provides a number of front panel error messages to assist in problem diagnosis. Table 8-1 lists these
messages and their meanings.

Table 8-1.

420

Error Messages

Error Message

Description

Solution

E A/D

A/D physical error

Call Rice Lake Weighing Systems (RLWS) Service.

EEEROM

EEPROM physical error

EVIREE

Virgin EEPROM

Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.

EPCKSM

Parameter checksum error

EACKSM

A/D calibration checksum error

A/D converter requires recalibration. Call RLWS Service.

EFCKSM

Printer format checksum error

Call RLWS Service.

ELCKSM

Load cell calibration checksum error

Recalibrate load cells.

EIDATA

Internal RAM checksum error

Call RLWS Service.

E REF

A/D reference error

A/D converter requires recalibration. Call RLWS Service.

ERROR

Internal program error

Check configuration. Run XE command (see Section

8.1.2 on

page

41) to determine error type.

Call RLWS Service if unable to clear error by cycling power or if error
recurs.

OVERFL

Overflow error

Weight value too large to be displayed.

------

Gross > overload limit

Gross value exceeds overload limit. Check configuration or signal
input level. Overload can be caused by input signal > 45 mV or
common mode voltage > 950 mV.

______

A/D underrange

A/D reading < –4 mV. Check scale for binding or damage.

RNGERR

GRADS > 100,000
WVAL > 100,000

Only shows up in Config mode.

NO COM

No communication with j-box

Check the cable or wireless connections or install a charged battery.

ERRSS

Serial Scale Error

Generic error. Need to troubleshoot system or call RLWS service.

----

Forks not level

Position forks within three degrees of level.

Wait

J-Box is powering up

This display is a normal action of the indicator.

LowBat

J-Box battery is getting low or dead.
May not occur if completely dead.

Change battery.

ErrCell

J-Box is reporting a generic error

This is a declared error and should look at test mode to diagnose the
problem.