24

CLS-420 Installation Manual

5.2

Menu Structures and Parameter Descriptions

The following sections provide graphic representations of the 420 menu structures. In the actual menu structure,
the setting you choose under each parameter are arranged horizontally. To save page space, menu choices are
shown in vertical columns. The factory default setting appears at the top of each column in bold type.
Most menu diagrams are accompanied by a table that describes all parameters and parameter values associated
with that menu. Default parameter values are shown in bold type.

5.2.1

Configuration Menu

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

GRADS

10000

0

number

ZTRKBN

ZRANGE

1.900000

MOTBAN

FS+2%

FS+9D

FS+1D

FS

OVRLOA

1

4

2

8

DIGFL1

32

16

64

number

1

4

2

8

DIGFL2

32

16

64

1

4

2

8

DIGFL3

32

16

64

BOTH

PBTARE

NOT ARE

KEYED

T AREFN

ALGOUT

DIG IN

SETPNT

PROGRM PFORMT

SERIAL

CALIBR

CONFIG FORMAT

8OUT

32OUT

16OUT

64OUT

DFSENS

DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

15HZ

60HZ

30HZ

7.5HZ

SMPRAT

VERS

number

1

number

Figure 5-4. Configuration Menu

CONFIG Menu

Parameter

Choices

Description

Level 2 submenus

GRADS

1000
number

Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1-100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate GRADS, use the formula, GRADS=Capacity / Display Divisions

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBND

0
number

Automatically zeroes the scale when within the range specified, as long as the input is
within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.

NOTE: For scales using linear calibration, do not set the zero tracking band to a value
greater than that specified for the first linearization point.

Table 5-2. Configuration Menu Parameters