9 analog output menu – Rice Lake CLS-420 Cargo Lift Scale Installation Manual User Manual

Page 46

Advertising
background image

36

CLS-420 Installation Manual

5.2.9

Analog Output Menu

The ALGOUT menu is shown only if the analog output option is installed. If the analog output option is installed,
configure all other indicator functions and calibrate the indicator itself before configuring the analog output.

XXXXXXX

XXXXXXX

ALGOUT

DIGIN

SETPNT

XXXXXXX

PROGRM

PFORMT

SERIAL

CALIBR

XXXXXXX

CONFIG

FORMAT

SOURCE

GROSS

0%

20%

OFFSET

ERRACT

FULLSC

HOLD

000000

number

MIN

10000

number

MAX

TWZERO

NET

TWSPAN

ZEROSC

VERS

MINNEG

OFF

ON

MAXNEG

OFF

ON

Figure 5-13. Analog Output Menu

Table 5-10. Analog Output Menu Parameters

ALG OUT Menu

Parameter

Choices

Description

Level 2 submenus

SOURCE

GROSS
NET

Specifies the scale tracked by the analog output.

OFFSET

0%
20%

Zero offset. Select 0% for 0–10 V or 0–20 mA output; select 20% for 4–20 mA output. This parameter
must be set before calibrating the analog output.

ERRACT

FULLSC
HOLD
ZEROSC

Error action. Specifies how the analog output responds to system error conditions.
Possible values are:

FULLSC: Set to full value (10 V or 20 mA)
HOLD: Hold current value
ZEROSC: Set to zero value (0 V or 4 mA)

MIN

000000
0–9999999

Specifies the minimum weight value tracked by the analog output. Specify a value in the range
0–9999999.

MINNEG

OFF
ON

Turn ON or OFF.

MAX

10000
0–9999999

Specifies the maximum weight value tracked by the analog output. Specify a value in the range
0–9999999

MAXNEG

OFF
ON

Turn ON or OFF.

TWZERO

_

Tweak zero. Enter tweak value to adjust the analog output zero calibration. Use a multimeter to monitor
the analog output value.

TWSPAN

_

Tweak span. Enter tweak value to adjust the analog output span calibration. Use a multimeter to monitor
the analog output value.