Rice Lake Digital/Diagnostic - iQube User Manual

Page 26

Advertising
background image

920i Indicator Configuration

23

SERIAL Menu (

iQUBE Submenu)

Parameter

Choices

Description

CONFIG

8/0/0/0
8/4/0/0
8/4/4/0
8/8/0/0
4/0/0/0
4/4/0/0
4/4/4/0
4/4/4/4
4/8/0/0

Specifies the

iQUBE

system configuration.

The first value (8 or 4) designates the number of channels on the

iQUBE

primary board;

the remaining three values represent attached secondary units, if any, and the number of
channels on each secondary board.

NOTE: The

iQUBE

allocates either 4 or 8 channels per board. Total channels cannot

exceed 16; an 8-channel board cannot be defined as a 4-channel board.

CELLS

CELLA1–CELLA8
List of cells for
CONFIG type

Depending on the value specified for CONFIG, load cells are listed using default names
with the format CELLxy, where x is the board identifier (A is the primary unit; B, C, and D
are secondary units), and y is a designation (1–4 or 1–8, depending on the number of
channels supported by the board), for the individual load cell.

Subparameters allow load cell ID, capacity, sensitivity, and serial number to be specified.
See descriptions under Level 4 Submenus.

SECTION

PAIRED
CIRCULAR

Specifies whether the load cells defined for the section are referenced to each other as
pairs or in a circular relationship. Sections with an odd number of cells must specify
CIRCULAR. Full diagnostic capability requires PAIRED section configuration.

See Section 2.3 on page 3 for information about load cell build order.

PLATFM

List of cells and
platforms

Use the Add softkey to associate load cells with platforms.

SYSTEM

List of platforms
and systems

Use the Add softkey to associate platforms with an

iQUBE

system. A system can

consist of 1–4 platforms.

ALGOUT

SOURCE
FULLSC

Specifies the source system and full scale value for an

iQUBE

analog output. See

descriptions under Level 4 Submenus.

DIGIO

BIT 1 – BIT 8

For

iQUBE

models with digital I/O, specifies the function of each of the digital I/O bits.

See the

920i

Installation Manual, PN 67887, for information about digital I/O functions.

IQDIAG

ZEROREF
CELLBAL
LDRIFT
PKtoPK
CELLEMU
MOTBAND

See Level 4 Submenus, Figure 3-3 on page 25, and Table 3-2 on page 26.

TESTCOM

VALIDAT

Select VALIDAT to run a communication validation test for one or more attached

iQUBE

s. The validation test returns the serial numbers of all attached

iQUBE

s.

PORTTYPE

RS232
RS485

Specifies the port designation.

Level 4 Submenus

ID

load_cell_ID

Specifies the load cell ID. Press enter to exit this parameter.

CAPCTY

75000
capacity

Enter the rated load cell capacity.

SENS

3.000000
sensitivity (mV/V)

Enter the actual or nominal load cell millivolt/volt sensitivity.

SRLNUM

serial_number

Optional. Enter the load cell serial number.

Table 3-1. Serial Menu Parameters,

iQUBE Submenu