6

iQUBE

Installation Manual

DIP switches on the core module of each

iQUBE

must

be set to indicate whether the unit is a primary or
secondary, the communication protocol and port used,
and the secondary address.

For the primary unit, DIP switches 1–3 must
all be set OFF. Set switches 5 and 6 based on
the protocol (RS-232 or RS-485) and port (J7
or J6) used to communicate with the host
device.

For secondary units, DIP switch 5 must be set
ON; use switches 1–3 to set the secondary
address as shown in Table 2-5.

See Section 2.9 on page 10 for more information
about the core module DIP switches.

Figure 2-3. 920i Wiring

DIP

Switch

Primary Unit

Secondary Unit

1–3

OFF, OFF, OFF

ON, OFF, OFF = SECONDARY1
OFF, ON, OFF = SECONDARY2
ON, ON, OFF = SECONDARY3

5

RS-232 = OFF

RS-485 = ON

ON

6

OFF = J7

ON = J6

iQUBE

Port

Table 2-5. Core Module Communications DIP Switches

J7

GND

S

HLD

YB

Z/TXD

A

/RXD

1

PRIMARY

iQUBE UNIT

iQUBE UNIT

TXD

RXD

GND

RS-232

J7

GND

S

HLD

YB

Z/TXD

A/RXD

1

PRIMARY

iQUBE UNIT

iQUBE UNIT

A R

S

-485 TXD +

A R

S

-485 TXD -

A R

S

-485 RXD +

RS-485

A R

S

-485 RXD -

120

120

RS-232 to

Host PC (920i)

RS-485 to

920i Serial

Option Board