0 appendix, 1 troubleshooting, 1 general steps – Rice Lake Digital/Diagnostic - iQube User Manual

Page 48: Appendix, General steps

Advertising
background image

Appendix

45

9.0

Appendix

9.1

Troubleshooting

The following section provides information to assist
for troubleshooting or repairing the

iQUBE

.

Materials needed to accomplish these steps are:

Revolution III

loaded onto a laptop PC. If using

Virtui

,

iRev

must be loaded on the PC

Digital voltmeter

9.1.1

General Steps

To determine that the

iQUBE

has power, check LED

D1 on the core module and LEDs for the load cells
assigned on each connector board (see Figure 2-2 on
page 4
). If no load cell LEDs (D18–D25) are on,
download configuration to the

iQUBE

.

Use a voltmeter to measure the voltage at connector
J10. Pin 1 should show 6 volts.

Verify communication LEDs D5 and D6 on the core
module are flashing (see Figure 2-2 on page 4). If
both LEDs are not flashing, check wiring and setup.

For

Virtui

applications, if the power and the LEDs are

all properly lit, use the

Diagnostic

softkey to go into

diagnostic mode. (This feature can be enabled in the
Feature menu in Config mode.) Use the diagnostic
displays to view the outputs of each load cell.

If no

Virtui

is available, connect the PC running

Revolution III

to connector J7 with switch 6 in the off

position. After connecting, upload the

iQUBE

c o n f i g u r a t i o n t o t h e P C t o v e r i f y t h e p r o p e r
configuration is still loaded. If not, download the
correct configuration to the

iQUBE

and recalibrate.

Also verify that the

iQUBE

software is the latest

version.

If the configuration looks good, check the outputs of
each load cell. Cell outputs should change when a
weight on top of each cell or when the load cell cable
is disconnected.

Use the interactive menu of

Revolution III

to monitor

the cells, platforms, or systems. If any of the load cells
reads incorrectly, correct the problem then test the
total system again.

If these steps does not resolve the problem, contact
Rice Lake Weighing Systems technical support for
further assistance.

Symptom

Cause/Remedy

iQUBE

seems dead

Verify power indicator LED on connector board is green.

Cycle power by pulling the

iQUBE

fuse and watch D1 on

core module cycle through red, green, then off. If it does not,
turn all switches off and only turn on 8. Cycle power again.
This will clear the core module.

No LEDs lit for load cells

Download from

Virtui

required after setup is completed.

Some LEDs lit green or red

Green light means good config and good cell; red indicates a
load cell fault or bad A/D conversion. Swap load cell cable to
another position.

No scale or

iQUBE

in

Virtui

configuration

Configuration setup and download needed to setup
configuration the

iQUBE

.

Ensure setup switch (DIP switch 4 on the core module) is ON
to enable download from

Virtui

to

iQUBE

.

Print button causes drift

Change print format in PFORMT to send print to a port not
connected to the

iQUBE

.

Slaves not configured

Cycle power on primary unit only. Secondary units will receive
new configuration data.

Table 9-1.

iQUBE Troubleshooting