0 diagnostics, 1 iqube diagnostic, 2 cell mate relationships – Rice Lake Digital/Diagnostic - iQube User Manual

Page 37: 3 diagnostic tests, Diagnostics, Lc 1

Advertising
background image

34

iQUBE

Installation Manual

6.0

Diagnostics

6.1

iQUBE Diagnostic

The

iQUBE

connector board provides bi-color cell

status LEDs (see Figure 2-2 on page 4) that indicate
the status of connected load cells. Table 6-1 lists the
cell status indications shown by the LEDs.

6.2

Cell Mate Relationships

Cell mates to a specific load cell are defined as the
load cells that are horizontally or vertically adjacent to
each other. In the platform illustration below LC2 and
LC 3 are cell mates to LC1 and LC2, LC3, and LC6
are cell mates to LC4.

Figure 6-1. Cell Mate Indications

6.3

Diagnostic Tests

The

iQUBE

provides a number of diagnostic tests, including boundary, weighing, and system tests (see Table 6-4

on page 37 for a list of diagnostic error messages). The table below describes the tests that can be enabled or
disabled in the diagnostics of the

iQUBE

.

LED Color

meaning

Green

Load cell good

Red

Load cell error

Off

Disabled/no load cell connected

Table 6-1. LED Status Indications

LC 1

LC 1

LC 5

LC 5

LC 3

LC 3

LC 2

LC 2

LC 4

LC 4

LC 6

LC 6

Test

Definition

Description

920i Default

Revolution III/

iQUBE Default

Zero
Reference

Which cell is out of tolerance at
zero

threshold: the upper limit of the zero reading of the
cells.

0.125mv

0.200mv

Per Cent threshold: the percentage of zero reading of
adjacent cells.

25%

20%

Delay: time to wait until the next testing process.

900 seconds

900 seconds

Cell Balance

Which cell is out of balance with its
section cell mate when weight is on
the scale.

Threshold

0.200mv

200mv

Per Cent tolerance

25%

20%

Load: minimum load needed to initiate the test

5000lb

5000lb

Load Cell Drift Checks for load cell outputs that

have changed over time with load
applied..

Limit

0.025mv

0.025mv

Per Cent limit

10%

10%

Time

30 seconds

30 seconds

Load: minimum load needed to initiate the test

1000lb

1000lb

Peak to Peak
Limit

Check for noisy load cell output not
attributed to environmental
conditions.

Limit

0.125mv

0.250mv

Motion Band

Defined value in Grads that defines
a scale not in motion

Motion band range: should be the same as
MOTBAND for truck scale

10lb

10lb

Cell Emulation Process of inputting a value based

on the matrix offset values gatered
at Calmatch time.

Type 1(standard truck scale): Auto 1, Manual 1
Type 2 (hopper scale configuration): Auto 1, Manual 2

Off

Off

Table 6-2.

iQUBE Diagnostic Settings