Table 6-4, Table 6-4 lists the system and cell error messages – Rice Lake Digital/Diagnostic - iQube User Manual

Page 40

Advertising
background image

Diagnostics

37

Table 6-4 lists the system and cell error messages.

920i Diagnostic Display Error Codes

Table 6-5 lists error codes that may be shown in the status area of the

920i

diagnostic display.Error codes are

shown in the format

Cell Error X

, where

X

is the code listed below. If the cell is reporting no error, the message

Cell

OK

is shown.

System Error Message

Cell Error Message

Error Categories

Boundary Error

Error 1

Cell Not Detected

AtoD Not Responding

Cell At Rail

Cell Over/Under Range

Weighing Error

Error 2

Cell Long Term Drift Error

Cell Balance Error

Cell Noise Error

Cell At Load Drift Error

Zero Reference Cell Error

Master (Primary Unit) Error

Error 3

Slave (Secondary) Comm Error

Temperature Error

Low Excitation Voltage

High Excitation Voltage

Low Power Supply

High Power Supply

DtoA1 Calibration Error

DtoA2 Calibration Error

Table 6-4. System and Cell Error Messages

920i

Error Code

Error Category

A

A-to-D not responding

B

Cell balance error

D

Load drift error

M

Cell not detected

N

Cell noise error

O

Cell over/under range error

R

Cell railed error

Z

Zero reference drift error

Table 6-5.

920i Diagnostic Display Error Codes